China Journal of Oral and Maxillofacial Surgery ›› 2016, Vol. 14 ›› Issue (5): 397-403.

• Original Articles • Previous Articles     Next Articles

Osseointegration study of SLM porous multi-rooted implants in rabbits

ZHANG Qing1, OU Yue-jian1, YOU Jia2, XU Hai-yan1, SUN Qi-jun1   

  1. 1.Department of Stomatology, Huzhou Central Hospital, Huzhou Hospital of Zhejiang University. Huzhou 313000;
    2. Key Laboratory of E&M, Ministry of Education&Zhejiang Province, Zhejiang University of Technology. Hangzhou 310032, Zhejiang Province, China
  • Received:2016-06-15 Online:2016-09-20 Published:2016-10-14

Abstract: PURPOSE: The aim of this study was to develop a multi-rooted dental implant (MRI) with a connected porous surface structure to facilitate osseointegration. METHODS: MRIs manufactured by selective laser melting (SLM) and commercial implants with resorbable blasting media (RBM)-treated surfaces were inserted into the hind limbs of New Zealand white rabbits. Osseointegration was evaluated periodically over 12 weeks by micro-computerized tomography (CT) scanning, histological analysis, mechanical push-out tests, and torsion tests. The results were analyzed by using SPSS19.0 software package for ANOVA intra-group and un-paired t test. RESULTS: Bone volume densities were consistently higher in MRI group than in RBM group throughout the study period, ultimately resulting in a peak value of 48.41% for MRI. Histological analysis revealed denser surrounding bone growth in MRIs; after 4 and 8 weeks, bone tissue had grown into the pore structures and root bifurcation areas, respectively. Biomechanics tests indicated binding of the porous MRIs to the neobone tissues, as push-out forces strengthened from 294.7 N to 446.5 N and maximum mean torsion forces improved from 81.2 N to 289.6 N(MRI), versus 34.8 N to 87.8 N in the RBM group. CONCLUSIONS: MRIs manufactured by SLM possess a connected porous surface structure that improves the osteogenic characteristics of the implant surface and may promote implant survival.

Key words: Implant design, Biomechanics, Selective laser melting, Multi-rooted implant, Osseointegration

CLC Number: